Table 14. Word B Generator
Mode Repeat

Calibration Procedure For Transponder Test Set Model An/Upm-155 Page Navigation
  19    20    21    22    23  24  25    26    27    28    29  
img
. . .
TB 9-6625-2337-24
Table 15. Word C Generator
TI
Oscilloscope
time indication
Test description
From pulse
To pulse
From pulse
WORD C 1 TO 2
1
2
1.93
2.07
WORD C 1 TO 3
1
3
3.93
4.07
WORD C 1 TO 4
1
4
5.93
6.07
WORD C 1 TO 5
1
5
9.93
10.07
WORD C 1 TO 6
1
6
11.93
12.07
WORD C 1 TO 7
1
7
13.93
14.07
WORD C 1 TO 8
1
8
15.93
16.07
WORD C 1 TO 9
1
9
17.93
18.07
WORD C 1 TO 10
1
10
19.93
20.07
WORD C 1 TO 11
1
11
21.93
22.07
WORD C 1 TO 12
1
12
23.93
24.07
WORD C 1 TO 13
1
13
25.93
26.07
WORD C 1 TO 14
1
14
27.93
28.07
WORD C 1 TO 15
1
15
29.93
30.07
WORD C 1 TO 16
1
16
31.93
32.07
WORD C 1 TO 17
1
17
33.93
34.07
WORD C 1 TO 18
1
18
35.93
36.07
WORD C 1 TO 19
1
19
37.93
38.07
WORD C 1 TO 20
1
20
39.93
40.07
WORD C 1 TO 21
1
21
41.93
42.07
WORD C 1 TO 22
1
22
43.93
44.07
WORD C 1 TO 23
1
23
45.93
46.07
WORD C 1 TO 24
1
24
47.93
48.07
WORD C 1 TO 25
1
25
49.93
50.07
WORD C 1 TO 26
1
26
51.93
52.07
WORD C 1 TO 27
1
27
53.93
54.07
WORD C 1 TO 28
1
28
55.93
56.07
WORD C 1 TO 29
1
29
57.93
58.07
WORD C 1 TO 30
1
30
59.93
60.07
WORD C 1 TO 31
1
31
61.93
62.07
WORD C 1 TO 32
1
32
63.93
64.07
WORD C 1 TO 33
1
33
65.93
66.07
WORD C 1 TO 34
1
34
67.93
68.07
WORD C 1 TO 35
1
35
69.93
70.07
WORD C 1 TO 36
1
36
71.93
72.07
(13) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
11. Mode Repeat
a. Performance Check
(1) Connect the pulse generator 9215 OUTPUT to TI TRIGGERS EXT IN (DO
NOT USE A 50  FEEDTHROUGH TERMINATION).
(2) Connect TI IFF VIDEO OUT CHAL/TAG to the oscilloscope channel 1 input
using a 75  feedthrough termination.