TB 9-6625-2337-24
(3) Connect TI TRIGGERS M4 GTC OUT to the oscilloscope channel 2 input using
a 75
feedthrough termination.
NOTE
Wait while pulse generator calibrates. This will take about 80
seconds.
(4) Reset and calibrate the pulse generator.
(5) Press TI keys as listed in (a) through (x) below:
(a) Press FUNC and ENTR keys.
(b) Use the arrow keys to highlight TRANSPONDER MENUS and press
ENTR key.
(c) 2, ENTR (CHALLENGES).
(d) Use the arrow keys to highlight M1 : ENTR (ON).
(e) Use the arrow keys to highlight M2: ENTR (ON).
(f) Use the arrow keys to highlight M3/A: ENTR (ON).
(g) Use the arrow keys to highlight MC: ENTR (ON).
(h) Use the arrow keys to highlight P1: ENTR (ON).
(i) Use the arrow keys to highlight P2: ENTR (ON).
(j) Use the arrow keys to highlight P3: ENTR (OFF).
(k) Use the arrow keys to highlight M4: SYNC.
(l) Use the arrow keys to highlight (S1: ON).
(m) Use the arrow keys to highlight (S2: ON).
(n) Use the arrow keys to highlight (S3: ON).
(o) Use the arrow keys to highlight (S4: ON).
(p) Use the arrow keys to highlight MODE REPEAT:.
(q) 8, ENTR.
(r) Press FUNC and keys.
(s) 5, ENTR (TIMING).
(t) Use the arrow keys to highlight CHALLENGE VIDEO:.
(u) 5 and press ENTR key.
(v) Press FUNC and keys.
(w) 14, ENTR (PRF/PRI).
(x) Use the arrow keys to highlight 0 TRIGGER: EXTERNAL +.
(6) Set the pulse generator to produce a 5 s wide, 5 V pulse with a period of 100 ms
in single step mode, (one pulse will be output each time the pulse generator is triggered).
(7) Set oscilloscope to trigger on channel 1, measurement type to Delta time, and to
measure the delay between channel 1 and channel 2.
(8) Trigger the pulse generator one time.
(9) Adjust the oscilloscope cursors to measure the delta time between the first pulse
on channel 1 and the channel 2 pulse. Oscilloscope will indicate within limits listed in first
(10) Count pulses displayed on channel 1. The count will be within limits listed in
25