Table 19. 7777 Reply Code
1st SIF Reply Video - Continued

Calibration Procedure For Transponder Test Set Model An/Upm-155 Page Navigation
  25    26    27    28    29  30  31    32    33    34    35  
img
. . .
TB 9-6625-2337-24
Table 20. VAR EMERG Codes - Continued
TI
Oscilloscope
Pulses to be
Minimum
Maximum
( s)
( s)
Test description
Settings
measured
VAR EMERG 1 - 5
----------------------------------------
1 to 5
49.20
49.40
VAR EMERG 1 - 6
----------------------------------------
1 to 6
69.50
69.70
VAR EMERG 1 - 7
----------------------------------------
1 to 7
73.85
74.05
VAR EMERG 1 - 8
----------------------------------------
1 to 8
94.15
94.35
+1 VAR EMERG 1 - 3
Use the arrow keys to highlight
1 to 3
25.50
25.80
VAR EMG:
Press ENTR to select +1
+1 VAR EMERG 1 - 4
----------------------------------------
1 to 4
45.80
46.10
+1 VAR EMERG 1 - 5
----------------------------------------
1 to 5
51.15
51.45
+1 VAR EMERG 1 - 6
----------------------------------------
1 to 6
71.45
71.75
+1 VAR EMERG 1 - 7
----------------------------------------
1 to 7
76.80
77.10
+1 VAR EMERG 1 - 8
----------------------------------------
1 to 8
97.10
97.40
-1 VAR EMERG 1 - 3
Use the arrow keys to
highlight
1 to 3
23.50
23.80
VAR EMG:
Press ENTR to select 1
-1 VAR EMERG 1 - 4
----------------------------------------
1 to 4
43.80
44.10
-1 VAR EMERG 1 - 5
----------------------------------------
1 to 5
47.15
47.45
-1 VAR EMERG 1 - 6
----------------------------------------
1 to 6
67.45
67.75
-1 VAR EMERG 1 - 7
----------------------------------------
1 to 7
70.80
71.10
-1 VAR EMERG 1 - 8
----------------------------------------
1 to 8
91.20
91.40
(19) Press TI keys as listed in (a) through (d) below:
Use the arrow keys to highlight REPLY SIGNAL: ENTR, (ID OF POS).
(a)
Use the arrow keys to highlight M1: 7777(ON).
(b)
Use the arrow keys to highlight M2: 7777(OFF).
(c)
Use the arrow keys to highlight F2: (OFF).
(d)
(20) Verify the pulse count and delta time between pulse repeats on channel 1 of
oscilloscope are within limits indicated in row 1 of table 21.
(21) Repeat technique of (19) and (20) above for remaining rows of table 21.
Table 21. ID, SIF1, SIF2, Etc.
TI
Oscilloscope
Delta time indication
Number of
Test description
Settings
Minimum
Maximum
pulses
( s)
( s)
1ST REP ID OF POS
--------------------------------
26
24.55
24.75
2
141
1ST REP ID OF POS M2
24.55
24.75
Press FUNC,
, ENTR (CHALLENGES)
Use the arrow keys to
highlight M1: (OFF)
Use the arrow keys to
highlight M2: (ON)
3 ress FUNC,
P
, ENTR (1st REPLY)
Use the arrow keys to
highlight M1:7777 (OFF)
Use the arrow keys to
highlight M2:7777 (ON)
Measure delta time between first and fourteenth pulse.
1
30