TB 9-6625-2337-24
(9) Set oscilloscope to trigger on channel 1 and measurement type to the delta time
between the 1st and 2d pulses on channel 1.
(10) Verify that the oscilloscope indicates within limits of table 18.
(11) Repeat technique of (8) through (10) above using remaining setting and limits
TI
Test description
Settings
Pulses to be
Minimum
Maximum
( s)
( s)
measured
M1 1030 P1 to P2
----------------------------------------
1 to 2
1.43
1.47
M1 1030 P1 to P3
----------------------------------------
1 to 3
2.88
2.92
M1 1030 P1 to P4
----------------------------------------
1 to 4
4.33
4.37
2
M2 6040 P1 to P2
1 to 2
5.78
5.82
Press FUNC,
, ENTR (CHALLENGES).
Use the arrow keys to highlight
M1: (OFF)
Use the arrow keys to highlight
M2: (ON)
M2 6040 P1 to P3
----------------------------------------
1 to 3
7.23
7.27
M2 6040 P1 to P4
----------------------------------------
1 to 4
8.68
8.72
M3 0301 P1 to P2
Use the arrow keys to highlight
1 to 2
11.58
11.62
M2: (OFF)
Use the arrow keys to highlight
M3/A: (ON)
M3 0301 P1 to P3
----------------------------------------
1 to 3
13.03
13.07
M3 0301 P1 to P4
----------------------------------------
1 to 4
14.48
14.52
MC 0406 P1 to P2
Use the arrow keys to highlight
1 to 2
15.93
15.97
M3A: (OFF)
Use the arrow keys to highlight
MC: (ON)
MC 0406 P1 to P3
----------------------------------------
1 to 3
17.38
17.42
MC 0406 P1 to P4
----------------------------------------
1 to 4
18.83
18.87
(12) Press TI keys as listed in (a) through (j) below:
Press FUNC, .
(a)
3, ENTR (1ST REPLY).
(b)
Use the arrow keys to highlight M1: 7777 (ON).
(c)
Use the arrow keys to highlight M2: 7777 (ON).
(d)
Use the arrow keys to highlight M3/A: 7777 (ON).
(e)
Use the arrow keys to highlight MC: 7777 (ON).
(f)
Press FUNC, .
(g)
2, ENTR (CHALLENGES).
(h)
Use the arrow keys to highlight M2: (ON).
(i)
Use the arrow keys to highlight MC: (OFF).
(j)
(13) Verify oscilloscope display indicates as listed in table 19 first row.
(14) Repeat technique of (12) and (13) above for remaining rows in table 19.
Oscilloscope display will indicate as listed in table 19.
28