TB 9-6625-2337-35
TI
Measurement
Test description
Settings
type
Minimum
Maximum
1st REP M2 Rise
----------------------------------------
Rise
--------------------
100 ns
Time
1st REP M2 Fall
----------------------------------------
Fall
--------------------
150 ns
Time
1st REP M2
----------------------------------------
Overshoot
--------------------
5%
Overshoot
1st REP M2 .45 s
----------------------------------------
Width
420 ns
480 ns
Pulse Width
1st REP M2 2.2 V
----------------------------------------
Top
2.1 V
2.3 V
Level
Press FUNC and ↑, keys
1st REP M2 5 V
Top
4.8 V
5.2 V
Level
5, ENTR (TIMING VIDEO
LEVELS)
Use the arrow keys to highlight
1ST REPLY VIDEO:
5, ENTR
1st REP M2 0.1 V
.1, ENTR
Top
0.0 V
0.2 V
Level
10.13 s
10.17 s
1st REP SIF+X
2.2, ENTR
Period
Press FUNC and ↑, keys
Period
3, ENTR (1ST REPLY)
Use the arrow keys to highlight
REPLY SIGNAL:
ENTR (SIF + X)
(8) Press TI keys as listed in (a) through (p) below:
(a)
Press FUNC, ENTR keys.
(b)
Use the arrow keys to highlight TRANSPONDER MENUS and press
ENTR key.
(c)
2, ENTR (CHALLENGES).
(d)
Use the arrow keys to highlight
M1: (ON).
(e)
Use the arrow keys to highlight
M2: (OFF).
(f)
Use the arrow keys to highlight
MC: (OFF).
(g)
Use the arrow keys to highlight
M3/A: (OFF).
(h)
Use the arrow keys to highlight
M4: (OFF).
Press FUNC, ↑.
(i)
(j)
3, ENTR (1ST REPLY).
(k)
Use the arrow keys to highlight
REPLY SIGNAL: ENTR, (SIF).
(l)
Use the arrow keys to highlight
M1: 1030 (ON).
(m)
Use the arrow keys to highlight
M2: 6040 (ON).
(n)
Use the arrow keys to highlight
M3/A: 0301 (ON).
(o)
Use the arrow keys to highlight
MC: 0406 (ON).
(p)
Use the arrow keys to highlight
F2: (ON).
(9) Set oscilloscope to trigger on channel 1 and measurement type to the delta time
between the 1st and 2d pulses on channel 1.
27