TB 9-6625-2337-35
(7) Verify that the oscilloscope delta time markers indicate within the limits listed
(8) Repeat technique of (6) and (7) for remaining rows in table 13.
TI
Minimum
Maximum
Test description
(s)
(s)
From pulse
To pulse
WORD A 1 TO 2
1
2
1.93
2.07
WORD A 1 TO 3
1
3
3.93
4.07
WORD A 1 TO 4
1
4
5.93
6.07
WORD A 1 TO 5
1
5
9.93
10.07
WORD A 1 TO 6
1
6
12.93
13.07
WORD A 1 TO 7
1
7
14.93
15.07
WORD A 1 TO 8
1
8
16.93
17.07
WORD A 1 TO 9
1
9
19.93
20.07
WORD A 1 TO 10
1
10
23.93
24.07
WORD A 1 TO 11
1
11
25.93
26.07
WORD A 1 TO 12
1
12
29.93
30.07
WORD A 1 TO 13
1
13
32.93
33.07
WORD A 1 TO 14
1
14
34.93
35.07
WORD A 1 TO 15
1
15
37.93
38.07
WORD A 1 TO 16
1
16
40.93
41.07
WORD A 1 TO 17
1
17
42.93
43.07
WORD A 1 TO 18
1
18
44.93
45.07
WORD A 1 TO 19
1
19
47.93
48.07
WORD A 1 TO 20
1
20
49.93
50.07
WORD A 1 TO 21
1
21
51.93
52.07
WORD A 1 TO 22
1
22
53.93
54.07
WORD A 1 TO 23
1
23
57.93
58.07
WORD A 1 TO 24
1
24
60.93
61.07
WORD A 1 TO 25
1
25
62.93
63.07
WORD A 1 TO 26
1
26
65.93
66.07
WORD A 1 TO 27
1
27
67.93
68.07
WORD A 1 TO 28
1
28
70.93
71.07
(9) Press TI keys as listed in (a) through (c) below:
(a) Use the arrow keys to highlight (M4:).
(b) ENTR.
(c) Use the arrow keys to highlight (WORD B).
(10) Repeat technique of (6) and (7) above for each row in table 14.
22