TB 9-6625-2337-35
Table 15. Word C Generator - Continued
Oscilloscope ∆ time indication
TI
Test description
From pulse
To pulse
From pulse
WORD C 1 TO 9
1
9
17.93
18.07
WORD C 1 TO 10
1
10
19.93
20.07
WORD C 1 TO 11
1
11
21.93
22.07
WORD C 1 TO 12
1
12
23.93
24.07
WORD C 1 TO 13
1
13
25.93
26.07
WORD C 1 TO 14
1
14
27.93
28.07
WORD C 1 TO 15
1
15
29.93
30.07
WORD C 1 TO 16
1
16
31.93
32.07
WORD C 1 TO 17
1
17
33.93
34.07
WORD C 1 TO 18
1
18
35.93
36.07
WORD C 1 TO 19
1
19
37.93
38.07
WORD C 1 TO 20
1
20
39.93
40.07
WORD C 1 TO 21
1
21
41.93
42.07
WORD C 1 TO 22
1
22
43.93
44.07
WORD C 1 TO 23
1
23
45.93
46.07
WORD C 1 TO 24
1
24
47.93
48.07
WORD C 1 TO 25
1
25
49.93
50.07
WORD C 1 TO 26
1
26
51.93
52.07
WORD C 1 TO 27
1
27
53.93
54.07
WORD C 1 TO 28
1
28
55.93
56.07
WORD C 1 TO 29
1
29
57.93
58.07
WORD C 1 TO 30
1
30
59.93
60.07
WORD C 1 TO 31
1
31
61.93
62.07
WORD C 1 TO 32
1
32
63.93
64.07
WORD C 1 TO 33
1
33
65.93
66.07
WORD C 1 TO 34
1
34
67.93
68.07
WORD C 1 TO 35
1
35
69.93
70.07
WORD C 1 TO 36
1
36
71.93
72.07
(13) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect the pulse generator 9215 OUTPUT to TI TRIGGERS EXT IN (DO
NOT USE A 50 Ω FEEDTHROUGH TERMINATION).
(2) Connect TI IFF VIDEO OUT CHAL/TAG to the oscilloscope channel 1 input
using a 75 Ω feedthrough termination.
(3) Connect TI TRIGGERS M4 GTC OUT to the oscilloscope channel 2 input using
a 75 Ω feedthrough termination.
NOTE
Wait while pulse generator calibrates. This will take about 80
seconds.
(4) Reset and calibrate the pulse generator.