TB 9-6625-2337-35
Table 34. KIR Characteristics - Continued
TI
Test description
Settings/connection
Measurement type
Minimum
Maximum
GTC TRIG
--------------------------------------
Overshoot
---------------
5
OVERSHOOT
Channel 2
GTC TRIG .8 S
--------------------------------------
Width
600 ns
1.0 s
PW
Channel 2
GTC TRIG DLY
--------------------------------------
Delta time channel 1
371.5 s
372.5 s
to channel 2
ISLS TRIG 4.5V
Move connection from TI KIR
V Top
3.5 V
5.5 V
LVL
GTC TRIG OUT to TI KIR
Channel 2
ISLS TRIG OUT.
ISLS TRIG RT
--------------------------------------
Rise Time
---------------
200 ns
Channel 2
ISLS TRIG FT
--------------------------------------
Fall Time
---------------
200 ns
Channel 2
ISLS TRIG
--------------------------------------
Overshoot
---------------
5%
OVERSHOOT
Channel 2
ISLS TRIG PW
--------------------------------------
Width
100 ns
1.5 s
Channel 2
CHVID - ISLS
Move the connection from the TI
Delta time channel 1
5.8 s
6.0 s
DLY
TRIGGERS M4 PRE OUT to
to channel 2
the KIR CH VID OUT.
(6) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect as follows:
(a) Connect TI VAR PULSE OUTPUT to pulse generator Ext Input using a
75 Ω feedthrough termination.
(b) Connect pulse generator (9211) OUTPUT to TI KIT EN TRIG IN using a
50 Ω feedthrough termination.
(c) Connect TI IFF VIDEO CHAL/TAG output to KIT CH VID IN using a
75 Ω feedthrough termination.
(d) Connect KIT REPLY OUT to oscilloscope channel 1 input using a 93 Ω
feedthrough termination.
(2) Press TI keys as listed in (a) through (g) below:
(a)
Press FUNC, ENTR keys.
(b)
Use the arrow keys to highlight GENERAL MENUS and press ENTR key.
(c)
2, ENTR (CHALLENGES).
(d)
Use the arrow keys to highlight M2:.
(e)
ENTR (OFF).
(f)
Use the arrow keys to highlight M4:.
(g)
ENTR, (WORD A).