TB 9-6625-2337-35
Table 35. KIT Characteristics - Continued
TI
Oscilloscope
Test description
Settings/connection
Measurement type
Minimum
Maximum
V top
3.5 V
5.5 V
KIT DISP 4.5V
Move connection from TI KIT
LVL
REPLY OUT to KIT DISP
OUT.
Press FUNC, ↑
2, ENTR (CHALLENGES)
Use the arrow keys to highlight
M4 NORMAL:
ENTR (DISP)
KIT DISP
--------------------------------------
Overshoot
--------
5%
OVERSHOOT
1.0 s
KIT DISP PW
--------------------------------------
Width
300 ns
197 s
200 s
DISP DELAY
Using a tee connector and an
Delta time from
WORD A
additional cable connect the
channel 2 to
TI VAR PULSE OUT to
channel 1
KIT EN TRIG IN.
18 s
20 s
DISP DELAY
Disconnect TI VAR PULSE
Delta time from
WORD B
channel 2 to
channel 2 input leaving
channel 1
connected to KIT EN
TRIG IN.
Using a tee connector
and an additional cable
connect the TI IFF VIDEO
OUT CHAL/TAG to
and KIT CH VID IN.
Use the arrow keys to highlight
M4:
ENTR (WORD B)
(6) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Press TI keys as listed in (a) through (r) below:
(a)
Press FUNC, ENTR keys.
(b)
Use the arrow keys to highlight GENERAL MENUS and press ENTR key.
(c)
2, ENTR (CHALLENGES).
(d)
Use the arrow keys to highlight M2:.
(e)
ENTR (OFF).
(f)
Use the arrow keys to highlight M4:.
(g)
ENTR, (WORD C).
Press FUNC, ↑.
(h)