• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION I. IDENTIFICATION AND DESCRIPTION

Calibration Procedure For Spectrum Analyzers An/Usm-677 (Model E4407B-H76) And Models E4407B, E4407B-E57, E4407B-H57 And E4407B-1D51Dr Page Navigation
1  2    3    4    5    6  
img
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
*TB 9-6625-2339-24
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
SPECTRUM ANALYZERS AN/USM-677
(AGILENT, MODEL E4407B-H76) AND
AGILENT MODELS E4407B, E4407B-E57,
E4407B-H57 AND E4407B-1D51DR
Headquarters, Department of the Army, Washington, DC
22 August 2011
Distribution Statement A: Approved for public release; distribution is unlimited.
REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
You can improve this manual. If you find any mistakes or if you know of a way to improve these
procedures, please let us know.  Mail your letter or DA Form 2028 (Recommended Changes to
Publications and Blank Forms) directly to: Commander, U.S. Army Aviation and Missile Command,
ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A reply will be furnished to
you.
You may also  send in  your comments electronically to our E-mail address:
2028@redstone.army.mil or by fax 256-842-6546/DSN 788-6546.  For the World Wide Web use:
https://amcom2028.redstone.army.mil. Instructions for sending an electronic 2028 can be found at
the back of this manual.
Paragraph
Page
SECTION
I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ............................
1
2
Forms, records, and reports .................................
2
2
Calibration description ........................................
3
2
II.
EQUIPMENT REQUIREMENTS
Equipment required .............................................
4
4
Accessories required.............................................
5
4
III.
CALIBRATION PROCESS
Preliminary instructions......................................
6
5
Equipment setup ..................................................
7
6
Frequency readout and marker frequency
8
8
count accuracy ...................................................
Frequency span accuracy .....................................
9
10
Noise sidebands ....................................................
10
11
Residual FM..........................................................
11
13
Input attenuator accuracy ...................................
12
15
Reference level accuracy ......................................
13
16
_________
*This bulletin supersedes TB 9-6625-2339-24, dated 21 July 2009, including all changes.