Figure 27. Frequency response connection 1.
Table 21. Frequency Response (4.0 GHz to 18 GHz)

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  46    47    48    49    50  51  52    53    54    55    56  
img
. . . . . .
TB 9-6625-2340-24
(33) Press TI MARKER PEAK SEARCH key.
The amplitude displayed will be
within limits specified in the first row table 19.
(34) Repeat the technique of (29) through (32) above for remaining settings in table 19.
The amplitude displayed will be within limits specified in table 19.
Table 19. Frequency Response (100 MHz to 2 GHz)
Signal generator
Test instrument
Readout amplitude
Adjusted output
Center
(dBm)
Frequency
level
frequency
Span
Min
Max
2 dBm
2.6
1.4
100 MHz
100 MHz
10 MHz
2 dBm
2.6
1.4
400 MHz
400 MHz
10 MHz
2 dBm
2.6
1.4
1 GHz
1 GHz
10 MHz
2 dBm
2.6
1.4
2 GHz
2 GHz
10 MHz
(35) Disconnect signal generator RF output connector from the power splitter input.
(36) Connect equipment as shown in figure 28.
Figure 28. Frequency response connection 2.
(37) Press TI keys as listed in (a) and (b) below:
(a)  CENTER FREQ, 2.4, GHz.
(b)  SPAN, 10 MHz.
(38) Set synthesized signal generator No. 1 frequency to 2.4 GHz and level output to
4 dBm.
(39) Set up measuring receiver to measure power at 2.4 GHz.
(40) Adjust synthesized signal generator No. 1 RF Output level for a measuring
receiver indication of 2.0 0.1 dBm.
51