Out of Band Response
Out of Band Response - Continued

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  60    61    62    63    64  65  66    67    68    69    70  
img
. . . . . . .
TB 9-6625-2340-24
(12) Repeat technique of (9) through (11) above for remaining frequencies listed in
table 36. The marker  amplitude indicated on TI display will be within limits specified
table 36.
Table 36. Out of Band Response (3 GHz)
Synthesized signal
generator No. 1
Test instrument
Displayed
Frequency
amplitude limit
(GHz)
(dB)
< 70
3.6428
< 60
6.3124
< 60
6.9642
(13) Press TI keys as listed in (a) through (c) below:
(a) SWEEP CONT.
(b) MARKER OFF.
(c) CENTER FREQUENCY, 6, GHz.
(14) Set synthesized signal generator No. 1 output frequency to 6.0 GHz.
(15) Press TI MARKER PRESEL PEAK key and wait for (PEAKING!) on TI
to disappear.
(16) Press TI MARKER OFF key.
(17) Adjust synthesized signal generator No. 1 output level to place the peak of the
signal trace at the top TI graticule line.
(18) Press TI keys as listed in (a) through (c) below:
(a) MARKER PEAK SEARCH.
(b) MARKER .
(c) SWEEP SINGLE.
(19) Change synthesized signal generator No. 1 output frequency to 2.5179 GHz.
(20) Press TI keys as listed in (a) and (b) below:
(a) SWEEP SINGLE.
(b) MARKER PEAK SEARCH.
(21) The marker  amplitude indicated on TI display will be within limits specified in
the first row of table 37.
(22) Repeat technique of (19) through (21) above for remaining frequencies listed in
table 37. The marker  amplitude indicated on TI display will be within limits specified
table 37.
Table 37. Out of Band Response (6 GHz)
Synthesized signal
generator No. 1
Test instrument
Displayed
Frequency
amplitude limit
(GHz)
(dB)
< 60
2.5179
< 60
3.1607
< 70
5.3572
(23) Press TI keys as listed in (a) through (c) below: