Out of Band Response - Continued
Out of Band Response - Continued

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  63    64    65    66    67  68  69    70    71    72    73  
img
. . . . . . .
TB 9-6625-2340-24
(b) MARKER PEAK SEARCH.
(51) The marker  amplitude indicated on TI display will be within limits specified in
the first row of table 40.
(52) Repeat technique of (49) through (51) above for remaining frequencies listed in
table 40. The marker  amplitude indicated on TI display will be within limits specified
table 40.
Table 40. Out of Band Response (15 GHz)
Synthesized signal
Test
generator No. 1
instrument
Displayed
Frequency
amplitude limit
(GHz)
(dB)
< 60
4.5715
< 60
5.2143
< 60
9.4643
< 60
10.1071
< 70
14.3572
(53) Press TI keys as listed in (a) through (c) below:
SWEEP CONT.
(a)
MARKER OFF.
(b)
CENTER FREQUENCY, 17, GHz.
(c)
(54) Set synthesized signal generator No. 1 output frequency to 17.0 GHz.
(55) Press TI MARKER PRESEL PEAK key and wait for (PEAKING!) on TI
to disappear.
(56) Press TI MARKER OFF key.
(57) Adjust synthesized signal generator No. 1 output level to place the peak of the
signal trace at the top TI graticule line.
(58) Press TI keys as listed in (a) through (c) below:
MARKER PEAK SEARCH.
(a)
MARKER .
(b)
SWEEP SINGLE.
(c)
(59) Change synthesized signal generator No. 1 output frequency to 5.2381 GHz.
(60) Press TI keys as listed in (a) and (b) below:
SWEEP SINGLE.
(a)
MARKER PEAK SEARCH.
(b)
(61) The marker  amplitude indicated on TI display will be within limits specified in
the first row of table 41.
(62) Repeat technique of (59) through (61) above for remaining frequencies listed in
table 41. The marker  amplitude indicated on TI display will be within limits specified
table 41.