Out of Band Response - Continued
Gain Compression

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  66    67    68    69    70  71  72    73    74    75    76  
img
. . . .
TB 9-6625-2340-24
(83) Repeat the technique used in (79) (d) through (g) and (80) through (82) above
using remaining TI and synthesized signal generator No. 1 settings in table 43.  The
marker  amplitudes indicated on TI display will be within limits specified in table 43.
Table 43. Multiple Responses
Synthesized
Test
signal generator
instrument
No. 1
Center frequency
Displayed
Frequency
setting
amplitude limit
(GHz)
(GHz)
(dB)
< 45
13.0
1.0679
< 45
13.0
1.91907
< 45
13.0
0.53395
(84) Press TI keys as listed in (a) through (g) below:
STP SIZE.
SHIFT, MARKER MKR/
(a)
SWEEP CONT.
(b)
MARKER NORMAL.
(c)
MARKER OFF.
(d)
FREQUENCY SPAN, 100, kHz.
(e)
CENTER FREQUENCY, 15, GHz.
(f)
REFERENCE LEVEL, 0, dBm.
(g)
(85) Set synthesized signal generator No. 1 output frequency to 15.0 GHz.
(86) Press TI MARKER PRESEL PEAK key and wait for (PEAKING!) on TI
to disappear.
(87) Press TI keys as listed in (a) through (d) below:
MARKER MKR CF.
(a)
MARKER NORMAL.
(b)
MARKER OFF.
(c)
FREQUENCY SPAN, 5, kHz.
(d)
(88) Adjust synthesized signal generator No. 1 output level to place the peak of the
signal trace at the top TI graticule line.
(89) Press TI keys as listed in (a) through (g) below:
MARKER NORMAL.
(a)
MARKER PEAK SEARCH.
(b)
MARKER .
(c)
SWEEP CONT.
(d)
CENTER FREQUENCY, 10.107133, GHz.
(e)
SHIFT, SWEEP CONT.
(f)
SWEEP SINGLE.
(g)
(90) When sweep is completed press TI SWEEP SINGLE key and wait for the sweep
to finish.
(91) Press TI MARKER PEAK SEARCH key.
(92) The marker
amplitude indicated on TI display will be within limits specified in
table 44.
71