TB 9-6625-2340-24
key.
(32) Press TI MARKER
(33) Change synthesized signal generator No. 1 output level to 9 dBm.
(34) Adjust synthesized signal generator No. 1 output level for a 15.00 0.02 dBm
indication on measuring receiver.
(35) Press TI REFERENCE LEVEL, 5, and dBm.
(36) Record the marker
amplitude indicated on TI display in displayed amplitude
Measuring receiver
Test
Indication
instrument
Level
Reference level
Displayed amplitude
(dBm)
(dBm)
(dBm)
15
5
(37) Press TI REFERENCE LEVEL, 15, and dBm keys.
(38) Adjust TI AMPLITUDE CAL control to place the signal trace 1 division below
the top graticule.
keys.
(39) Press TI MARKER NORMAL and
(40) Change synthesized signal generator No. 1 output level to 1 dBm.
(41) Adjust synthesized signal generator No. 1 output level for a 5.00 0.02 dBm
indication on measuring receiver.
(42) Press TI REFERENCE LEVEL, 5, and dBm.
(43) Record the marker amplitude indicated on TI display in Displayed Amplitude
Measuring receiver
Test
indication
instrument
Level
Reference level
Displayed amplitude
(dBm)
(dBm)
(dBm)
5
5
(44) Calculate gain compression using the formula below. Calculated value will be
Amplitude @ 5 dBm amplitude @ 15 dBm = gain compression
Calculated gain compression
(dBm)
Min
Max
1
1
(45) Press TI keys as listed in (a) through (h) below:
RES BW, 3, MHz.
(a)
VIDEO BW, 3, MHz.
(b)
75