TB 9-6625-2341-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of ScopeMeter, Fluke, Model 105B. The manufacturer's manual was used as the
prime data source in compiling these instructions. The equipment being calibrated will be
referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications1
Vertical:
Range: 5 mV/Div to 100 mV/Div
Accuracy: (2% setting +1 pixel)
-3 db down from dc to ≥100 MHz
≤10 MHz ≤1.4 div
Trigger sensitivity:
≤100 MHz ≤2.1 div
≤150 MHz ≤2.9 div
Timebase:
Range: 5 ns/div to 1 s/div
Accuracy: (0.08% +1 pixel)
Dc volts:
Range: 100 mV, 300 mV, 1, 3, 10, 30, 100, and 300 V
Accuracy: (0.5% +5 counts)
Ac volts:
Range: 100 mV, 300 mV, 1, 3, 10, 30, 100, and 250 V
Accuracy:
50 to 60 Hz
(1% +10 counts)
20 Hz to 20 kHz
(2% +15 counts)
5 Hz to 1 MHz
(3% +20 counts)
5 Hz to 5 MHz
(10% +25 counts)
Ext.mV dc volt:
Range: 300 mV, 3 V
Accuracy: (0.5% +5 counts)
Ext.mV ac volt:
Range: 300 mV, 3 V
Accuracy: 50 and 60 Hz, (2% +15 counts)
See footnote at end of table.