TB 9-6625-2341-35
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
OSCILLOSOPE
Range: .02 V p-p - 100 V p-p (square
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
wave)
(5820A-5C-GHz)
Accuracy: +/-0.75%
Time markers:
Range:
Accuracy:
1uSec
+/-0.01%
Range capability:
0 - 0.5 V p-p @ 150 MHz (Sine wave)
0 - 0.3 V p-p @ 10 MHz (Sine wave)
0 - 0.13 V p-p @ 100 MHz (Sine wave)
Range capability:
50 kHz 100 MHz @ 120 mV p-p (square
wave).
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d.
When indications specified in
paragraphs 8 through 15 are not within
tolerance,
through 15. Do not perform adjustments if all other parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.