TB 9-6625-2341-35
(4)
(F5) SELECT ITEM.
(5)
Toggle ▲ or ▼ to display/highlight OFF.
(6)
(F5) SELECT ITEM.
(7)
(F1) CLOSE.
(b)
"Input B" probe to 1:1 by pressing (1) through (8) below:
(1)
INPUT B.
(2)
(F1) MORE INPUT B.
(3)
Toggle ▲ or ▼ to display/highlight PROBE B MENU.
(4)
(F5) SELECT ITEM which should display/highlight PROBE on INPUT B.
(5)
(F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(6) Toggle ▲ or ▼ to display/highlight 1:1.
(7) (F5) SELECT ITEM.
(8) (F1) CLOSE which should display/highlight INPUT B.
(c) "Input B" to on by pressing (1) through (5).
(1) Toggle ▲ or ▼ to display/highlight INPUT B.
(2)
(F5) SELECT ITEM.
(3) Toggle ▲ or ▼ to display/highlight ON.
(4) (F5) SELECT ITEM.
(5) (F1) CLOSE.
(d) Auto to manual range by toggling (mV RANGE V) on right side,
under INPUT B pushbutton to display/highlight B 5mVDC 1:1.
(e) "Input B" Trigger source and manual level by pressing (1)
through (10) below:
(1) TRIGGER.
(2) (F1) MORE TRIGGER.
(3) Toggle ▲ or ▼ to display/highlight TRIGGER SOURCE.
(4) (F5) SELECT ITEM.
(5) Toggle ▲ or ▼ to display/highlight INPUT B.
(6) (F5) SELECT ITEM which should display/highlight TRIGGER SOURCE.
(7) (F1) CLOSE.
(8) (F5) ADJUST LEVEL.
(9) Toggle ▲ or ▼ to change trigger level starting with (LEVEL 0.000 V) display.
(11) Perform steps as listed in (a) through (c) below for each row in table 4.
(a) Under INPUT B pushbutton on right side, toggle (mV RANGE V) to
display TI Range settings.
(b) Set oscilloscope calibrator to Output voltage settings at 1 kHz while
adjusting edit amplitude output dial for 4 vertical divisions.
NOTE
If necessary, adjust TI Trigger level by toggling ▲ or ▼
pushbutton for a stable display.
8