TB 9-6625-2341-35
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle ▲ or ▼ to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle ▲ or ▼ to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all Scope Meter settings have been set to original setting then (F1)
CLOSE), which should display/highlight RESET SCOPEMETER.
(h) Toggle ▲ or ▼ to display/highlight RESET SCOPE.
(i) (F5) SELECT ITEM.
(j) (For original settings, SCOPE mode needs to be selected. Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all Scope Meter settings have been
set to original setting then (F1) CLOSE), which should display/highlight RESET SCOPE.
(k) (F1) CLOSE which should display/highlight RESET MENU.
(l) (F1) CLOSE which should display/highlight CONTR.
(3) Press pushbutton sequence as listed in (a) through (e) below:
(a) "Input A" to off by pressing (1) through (7) below:
(1) INPUT A.
(2) (F1) MORE INPUT A.
(3) Toggle ▲ or ▼ to display/highlight INPUT A.
(4) (F5) SELECT ITEM.
(5) Toggle ▲ or ▼ to display/highlight off.
(6) (F5) SELECT ITEM.
(7) (F1) CLOSE.
(b) "Input B" probe to 1:1 by pressing (1) through (9) below:
(1) INPUT B to display/highlight DISPLAY INPUT B.
(2) (F1) MORE INPUT B.
(3) Toggle ▲ or ▼ to display/highlight PROBE B MENU.
(4) (F5) SELECT ITEM which should display/highlight PROBE on INPUT B.
(5) (F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(6) Toggle ▲ or ▼ to display/highlight 1:1.
(7) (F5) SELECT ITEM.
(8) (F1) CLOSE which should display/highlight INPUT B.
(9) (F1) CLOSE which should display/highlight DISPLAY INPUT B.
(c) Auto to manual range by toggling (mV RANGE V) on right side, under
INPUT B pushbutton to display/highlight B 50mVDC 1:1.
(d) "Input B" Trigger source and Trigger slope by pressing (1) through (11) below:
(1) TRIGGER.
(2) (F1) MORE TRIGGER.
(3) Toggle ▲ or ▼ to display/highlight TRIGGER SOURCE.