TB 9-6625-2341-35
(6) (F5) SELECT ITEM.
(7) (F1) CLOSE.
(c) "Input B" probe to 1:1 by pressing (1) through (8) below:
(1) (F1) MORE INPUT B.
(2) Toggle ▲ or ▼ to display/highlight PROBE B MENU.
(3) (F5) SELECT ITEM which should display/highlight PROBE on INPUT B.
(4) (F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(5) Toggle ▲ or ▼ to display/highlight 1:1.
(6) (F5) SELECT ITEM.
(7) (F1) CLOSE which should display/highlight INPUT B.
(8) (F1) CLOSE which should display/highlight DISPLAY INPUT B.
(d) Auto to manual range by toggling (mV RANGE V) on right side, under
INPUT B pushbutton to display/highlight B 20mVDC 1:1.
(e) "Input B" Trigger source by pressing (1) through (7) below:
(1) TRIGGER.
(2) (F1) MORE TRIGGER.
(3) Toggle ▲ or ▼ to display/highlight TRIGGER SOURCE.
(4) (F5) SELECT ITEM.
(5) Toggle ▲ or ▼ to display/highlight INPUT B.
(6) (F5) SELECT ITEM.
(7) (F1) CLOSE.
(11) RESET oscilloscope calibrator.
(12) Set oscilloscope calibrator LEVSINE output to produce a 120 mV at 50 kHz and
adjust edit amplitude output dial for 6 vertical divisions on TI.
(13) Set oscilloscope calibrator frequency output to produce 80 MHz.
the amplitude down to 4.2 divisions on TI. If frequency is not greater than or equal to 100
MHz, perform b below.
(15) Set oscilloscope calibrator to standby.
b. Adjustments.
Complete adjustments are in
paragraph 16 (R).
12