TB 9-6625-2341-35
(6) Toggle ▲ or ▼ to display/highlight 1:1.
(7) (F5) SELECT ITEM.
(8) (F1) CLOSE which should display/highlight INPUT A.
(9) (F1) CLOSE which should display/highlight DISPLAY INPUT A.
(d) Auto to manual range by toggling (mV RANGE V) on left side, under
INPUT A pushbutton to display/highlight A 50mVDC 1:1.
(e) "Input A" Trigger source and Trigger slope by pressing (1) through (7) below:
(1)
TRIGGER.
(2)
(F1) MORE TRIGGER.
(3)
Toggle ▲ or ▼ to display/highlight TRIGGER SOURCE.
(4)
(F5) SELECT ITEM.
(5)
Toggle ▲ or ▼ to display/highlight INPUT A.
(6)
(F5) SELECT ITEM.
(7)
(F1) CLOSE.
(14) Set oscilloscope calibrator LEVSINE output to produce a .5 V p-p at 150 MHz.
(15) Using oscilloscope calibrator, adjust edit dial amplitude to 2.9 vertical
divisions on TI. If TI signal is not triggered, perform b below.
(16) Set oscilloscope calibrator LEVSINE output to produce a 0.13 V p-p at 100
MHz.
(17) Using oscilloscope calibrator, adjust edit dial amplitude to 2.1 vertical
divisions on TI. If TI signal is not triggered, perform b below.
(18) Set range by toggling (mV RANGE V) on left side, under INPUT A
pushbutton to display/highlight A 200 mV DC 1:1.
(19) Set oscilloscope calibrator LEVSINE output to produce a .3 V p-p at 10 MHz.
(20) Using oscilloscope calibrator, adjust edit dial amplitude to 1.4 vertical
divisions on TI. If TI signal is not triggered, perform b below.
b. Adjustments.
Complete adjustments are in
paragraph 16 (R).
Scope Time-base
a. Performance Check
(1) Reset by pressing pushbuttons sequence (a) through (l) below:
(a) USER OPTIONS.
(b) (F1) MORE OPTION.
(c) Toggle ▲ or ▼ to display/highlight RESET MENU.
(d) (F5) SELECT ITEM.
(e) Toggle ▲ or ▼ to display/highlight RESET SCOPEMETER.
(f) (F5) SELECT ITEM.
(g) (If all Scope Meter settings have been set to original setting then (F1)
CLOSE), which should display/highlight RESET SCOPEMETER.
(h) Toggle ▲ or ▼ to display/highlight RESET SCOPE.
(i) (F5) SELECT ITEM.
(j) (For original settings, SCOPE mode needs to be selected. Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all ScopeMeter settings have been
set to original setting then (F1) CLOSE), which should display/highlight RESET SCOPE.