TB 9-6625-2341-35
Calibrator
Test instrument
Voltage output
Range settings
Voltage indication
settings
(Hz)
(V rms ac)
(V rms ac)
(V ac)
Min
Max
.1
55
100 m
98.00 m
102.0 m
.3
55
300 m
296.0 m
304.0 m
1
20
1
0.965
1.035
1
55
1
0.980
1.020
1
19
k
1
0.965
1.035
1
21
k
1
0.950
1.050
1
500
k
1
0.950
1.050
1
900
k
1
0.950
1.050
3
55
3
2.960
3.040
10
55
10
9.800
10.20
30
55
30
29.60
30.40
100
55
100
98.00
102.0
200
55
300
197.0
203.0
(7) RESET calibrator.
(8) Connect equipment as shown in figure 6 below:
(9) Perform steps as listed in (a) through (d) below for each row in table 7.
(a) Under INPUT A pushbutton on left side, toggle (mV RANGE V) to
display 1 V rms AC.
20