TB 9-6625-2341-35
(b) Set calibrator wideband to 1 volt output at selected Frequency.
(c) If TI Voltage indications are not within specified limits, perform b below.
(d) Set calibrator to stand-by.
Calibrator
Test instrument
Frequency settings
Voltage indication
(MHz)
(V ac)
Min
Max
1.1
0.875
1.125
5
0.875
1.125
(10)Press pushbutton sequence as listed in (a) through (d) below:
(a) MEASURE MENU to display/highlight (V, Hz, dB, TIME).
(b) (F3) Hz.
(c) Toggle ▲ or ▼ to display/highlight Hz METER.
(d) (F5) SELECT ITEM.
(11) Set calibrator frequency to 1 MHz output.
(12) If TI frequency indication is not within 0.993 and 1.007 MHz, perform b below.
(13) Set calibrator to standby.
b.
Adjustments.
Complete adjustments are in
paragraph 16 (R).
a. Performance Check
(1)
Connect equipment as shown in figure 7 below: