TB 9-6625-2341-35
(c) Toggle ▲ or ▼ to display/highlight PROBE on INPUT A.
(d) (F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(e) Toggle ▲ or ▼ to display/highlight 1:1.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE which should display/highlight CONTR.
(4) RESET calibrator.
(5) Set measure to V DC by pressing (a) through (d) below:
(a) MEASURE MENU to display/highlight (V, Hz, dB, TIME).
(b) (F2) V.
(c) Toggle ▲ or ▼ to display/highlight V DC METER.
(d) (F5) SELECT ITEM.
(6) Select display change by pressing (a) through (h) below:
(a) SUB MENU.
(b) (F1) MORE METER.
(c) Toggle ▲ or ▼ to display/highlight READINGS on DISPLAY.
(d) (F5) SELECT ITEM which should display/highlight (1, 2, 3, 4).
(e) Toggle ▲ or ▼ to display/highlight 1.
(f) (F5) SELECT ITEM.
(g) (F1) CLOSE.
(h) SCOPEMETER/MAIN MENU to display SCOPE/METER/Ω/EXT.mV and etc.
(7) Perform steps as listed in (a) through (d) below for each row in table 5.
(a) Under INPUT A pushbutton left side, toggle (mV RANGE V) to display TI
Range settings.
(b) Set calibrator to Output voltage settings.
(c) If TI Voltage indications are not within specified limits, perform b below.
(d) Set calibrator to standby.
Test instrument
Output volt settings
Range settings
Voltage indication
(DC)
(DC)
(DC)
Min
Max
.1
V
100
mV
99.00 mV
101.00 mV
.3
V
300
mV
298.0
mV
302.0
mV
1
V
1
V
0.990 V
1.010 V
3
V
3
V
2.980 V
3.020 V
10
V
10
V
9.900 V
10.10 V
30
V
30
V
29.80 V
30.20 V
100
V
100
V
099.0
V
101.0
V
300
V
300
V
298.0
V
302.0
V
b.
Adjustments.
Complete adjustments are in
paragraph 16 (R).
18