TB 9-6625-2341-35
(k) (F1) CLOSE which should display/highlight RESET MENU.
(l) (F1) CLOSE which should display/highlight CONTR.
(2) RESET oscilloscope calibrator.
(3) Press pushbutton sequence as listed in (a) through (d) below:
(a) "Input B" to off by pressing (1) through (7) below:
(1) INPUT B.
(2) (F1) MORE INPUT B.
(3) Toggle ▲ or ▼ to display/highlight INPUT B.
(4) (F5) SELECT ITEM.
(5) Toggle ▲ or ▼ to display/highlight off.
(6) (F5) SELECT ITEM.
(7) (F1) CLOSE.
(b) "Input A" probe to 1:1 by pressing (1) through (9) below:
(1) INPUT A.
(2) (F1) MORE INPUT A.
(3) Toggle ▲ or ▼ to display/highlight PROBE A MENU.
(4) (F5) SELECT ITEM which should display/highlight PROBE on INPUT A.
(5) (F5) SELECT ITEM which should display/highlight (1:1, 10:1, 100:1,
1000:1, etc.).
(6) Toggle ▲ or ▼ to display/highlight 1:1.
(7) (F5) SELECT ITEM.
(8) (F1) CLOSE which should display/highlight INPUT A.
(9) (F1) CLOSE which should display/highlight DISPLAY INPUT A.
(c) Auto to manual range by toggling (mV RANGE V) on left side, under
INPUT A pushbutton to display/highlight A 500mVDC 1:1.
(d) Time range by toggling (s TIME ns) to read 1 s/Div.
(4) Set oscilloscope calibrator MARKER output to produce 1 s pulses.
(5) Toggling horizontal ◄ MOVE ► pushbutton to align 2d time marker with 2d
vertical graticule line.
NOTE
If necessary, for a stable trigger press TRIGGER, (F5)
ADJUST LEVEL, then toggle ▲ or ▼ pushbutton to change
trigger level starting with (LEVEL 0.000 V) display.
(6) If 10th marker is not within +/-0.04 of vertical graticule line, perform b below.
NOTE
0.04 division = 1 pixel
16