TB 9-6625-2341-35
(8) Connect TI Optical Interface (PM 9080, RS-232-C cable) to controller COM 2.
(9) Execute Refresh Utility program, read the screen, and follow the instructions.
NOTE
If the screen calls for +12 volts to complete this restoration,
apply the +12 volts from calibrator to TI.
(10)
(1) TI should be allow at least 20 minutes for warm-up.
(2) Complete reset as listed in (a) through (e) below:
(a) Select reset scope by pressing pushbutton sequence (1) through (10) below:
(1) USER OPTIONS.
(2) (F1) MORE OPTION.
(3) Toggle ▲ or ▼ to display/highlight RESET MENU.
(4) (F5) SELECT ITEM.
(5) Toggle ▲ or ▼ to display/highlight RESET SCOPEMETER.
(6) (F5) SELECT ITEM.
(7) (If all Scope Meter settings have been set to original setting then (F1)
CLOSE), which should display/highlight RESET SCOPEMETER.
(8) Toggle ▲ or ▼to display/highlight RESET SCOPE.
(9) (F5) SELECT ITEM.
(10) (For original settings, SCOPE mode needs to be selected. Change to
SCOPE mode now? (F1) YES then (F1) CLOSE) or (if all Scope Meter settings have
26