TB 9-6625-2341-35
(d) Oscilloscope calibrator to STBY.
(15) Select pulse response of *10 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼CAL 7).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 3 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(16) Select pulse response of *100 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 8).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(17) Select pulse response of *1000 attenuation by pressing pushbuttons sequence (a)
through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 9).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 50 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(18) Select gain for 5 mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 10).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(19) Select gain for 10mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 11).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 50 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(20) Select gain for 20mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 12).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 100 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(21) Select gain for 50mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 13).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 200 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(22) Select gain for 100mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼CAL 14).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 500 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(23) Select gain for 200mV by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 15).
29