TB 9-6625-2341-35
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 1 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(24)
Select gain for 2 V by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼to display/highlight (▲ or ▼CAL 16).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 10 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(25)
Select gain for 20V by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼to display/highlight (▲ or ▼ CAL 17).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 100 V p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(26)
Shift gain *1 mode by pressing pushbuttons sequence (a) through (d) below:
(a) Toggle ▲ or ▼to display/highlight (▲ or ▼ CAL 18).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 200 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Oscilloscope calibrator to STBY.
(27)
Shift gain /8 mode by pressing pushbuttons sequence (a) through (c) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 19).
(b) Set oscilloscope calibrator to produce a square wave, 1 kHz at 20 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(28)
Reset oscilloscope calibrator.
(29)
Connect equipment as shown in figure 11 below:
OSCILLOSCOPE CALIBRATOR
(30) Set channel A and B, 50 percent and 90 percent trigger level by pressing
pushbuttons sequence (a) through (i) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 20).
(b) Set calibrator to produce an output of 10 kHz at 707 mV p-p.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
30