TB 9-6625-2341-35
(n) Set calibrator to produce a 30 V dc signal.
(o) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(p) Set calibrator to STBY.
(q) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 24).
(r) Set calibrator to produce a 100V DC signal.
(s) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(t) Set calibrator to STBY.
(u) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 7).
(v) Set calibrator to produce a 300V DC signal.
(w) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(x) Reset calibrator.
(50) Select EXT mV, 300 mV, 3V, range/gain by pressing pushbuttons sequence (a)
through (g) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 8).
(b) Set calibrator to produce a 300 mV DC signal.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 9).
(f) Set calibrator to produce a 3 V dc signal.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(51) Reset calibrator.
(52) Select ohms 100, 1 k, 10 k, 100 k, 1 M, 10 M, range by pressing pushbuttons
sequence (a) through (w) below:
(a) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 11).
(b) Set calibrator two wire resistance to produce 100 ohms.
(c) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(d) Set calibrator to STBY.
(e) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 12).
(f) Set calibrator to produce 1 k ohms.
(g) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(h) Set calibrator to STBY.
(i) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 13).
(j) Set calibrator to produce 10 k ohms.
(k) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(l) Set calibrator to STBY.
(m) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 14).
(n) Set calibrator to produce COMP OFF 100 k ohms.
(o) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(p) Set calibrator to STBY.
(q) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 15).
(r) Set calibrator to produce 1 M ohms.
(s) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(t) Set calibrator to STBY.
(u) Toggle ▲ or ▼ to display/highlight (▲ or ▼ CAL 16).
(v) Set calibrator to produce 10 M ohms.
33