TB 9-6625-2341-35
(w) Press (F5) READY, then wait till display/highlight READY is not highlighted.
(53) Set calibrator to standby.
(54) Disconnect calibrator leads from TI.
(56) Set calibrator to produce +12 volts.
(57) Press (F3) METER.
(58) Store Cal and Generator out by pressing pushbuttons sequence (a) through (p) below:
(a) (F5) CAL STORE.
(b) Displayed is "New Calibration array will be saved in Flash memory. Are you sure?"
(c) (F1) YES.
(d) Displayed SCOPE/METER/Ω/EXT.mV and etc.
(e) Press SPECIAL FUNCT.
(f) (F2) GENERATE MENU.
(g) Toggle ▲ or ▼ to display/highlight (F2) GENERATOR OUTPUT: ON/OFF.
(h) (F5) SELECT ITEM to display/highlight GENERATOR OUTPUT: ON.
(i) (F1) CLOSE.
(j) SCOPE METER/MAIN MENU to display SCOPE/METER/Ω/EXT.mV and etc.
(k) Press pushbutton BLACKLIGHT (sun shaped pushbutton just below
SPECIAL FUNCT key) with AUTO SET together, then release.
(l) (F4) CALIBRATE SCOPEMETER.
(m) Displayed are the numbers of calibration memories free. Press (F1) CLOSE.
(n) (F3) METER.
(o) Set calibrator to standby.
(p) Remove + 12 volts from calibrator to TI.
(59) Connect equipment as shown in figure 13 below:
NOTE
(60) Select current ramp by pressing pushbuttons sequence (a) through (c) below:
34