TB 9-6625-2341-35
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications1
Range: 30, 300, 3 k, 30 k, 300 k, 3 M, and 30 M ohm
Accuracy: 30 ohm, (2.5% +25 counts), 300 ohm to 30 M ohm, (0.5% + 5
counts)
Range: 1 Hz to 5 MHz
Accuracy: (0.5% +2 counts)
1
All specifications are +/-(% of indication +counts).
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and the TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: 50 Ω feedthrough, BNC plug to BNC jack, Tektronix, Type 011-0049-01(011-
0049-01).
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
CALIBRATOR
Range: 100 mV ac 200 V ac
Fluke, Model 5720A (5700A/EP)
Accuracy:
(p/o MIS-35947); w amplifier,
5 Hz - 20 Hz
+/-1.25%
Fluke 5725A/AR (5725A/AR)
20 Hz - 50 Hz
+/-0.875%
50 Hz - 60 Hz
+/-1.25%
60 Hz 20 kHz
+/-0.875%
20 kHz 1 MHz
+/-1.25%
1 MHz 5 MHz
+/-3.125%
Volt range:
Accuracy:
.1 - 300 V dc
+/-0.166%
Ohm range:
Accuracy:
10 ohm
+/-
1.25%
100 - 10 M ohm
+/-0.25%
Accuracy:
1 MHz
+/-0.175%