TB 9-6625-2346-35
Oscilloscope calibrator
Test instrument
MARKER
Err display
A AND B SEC/DIV
Linearity over any 2 of
output
limit
setting
center 8 divisions
(%)
(div)
.05 s
≤ 0.1
10 nS/D
3
(14) Set TI A AND B SEC/DIV to .1 s.
(15) Ensure oscilloscope calibrator is set for a CHAN 1, MARKER mode output
of 10 ns/div.
(16) Adjust TI A TRIGGER LEVEL, A INTENSITY, and CH 1 ⇕POSITION
controls for suitable viewing.
(17) Adjust TI ⇐POSITION⇒ control to align 1st time marker that is 25 ns beyond
start of sweep with 2nd vertical graticule line.
(18) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
align 10th time marker with 10th vertical graticule line. If oscilloscope calibrator err display
and TI linearity are not within limits specified in first row table 11, perform b (18) through
(21) below.
Oscilloscope calibrator
Test instrument
MARKER
Err display
A AND B SEC/DIV
Linearity over any 2 of
output
limit
setting
center 8 divisions
(%)
(div)
.1 s
≤ 0.1
10 nS/D
3
20 nS/D
.2 s
≤ 0.1
3
50 nS/D
.5 s
≤ 0.1
3
.1 S/D
1
s
≤ 0.1
3
.2 S/D
2
s
≤ 0.1
3
.5 S/D
5
s
≤ 0.1
3
1 S/D
10
s
≤ 0.1
3
2 S/D
20
s
≤ 0.1
3
5 S/D
50
s
≤ 0.1
3
10 S/D
.1 ms
≤ 0.1
3
20 S/D
.2 ms
≤ 0.1
3
50 S/D
.5 ms
≤ 0.1
3
.1 mS/D
1
ms
≤ 0.1
3
.2 mS/D
2
ms
≤ 0.1
3
.5 mS/D
5
ms
≤ 0.1
3
1 mS/D
10
ms
≤ 0.1
3
2 mS/D
20
ms
≤ 0.1
3
5 mS/D
50
ms
≤ 0.1
3
≤ 0.1
10 mS/D
3
.1 sec
A ONLY
≤ 0.1
20 mS/D
3
.2 sec
A ONLY
≤ 0.1
50 mS/D
3
.5 sec
A ONLY
17