TB 9-6625-2346-35
Oscilloscope calibrator
Test instrument
MARKER
Err display
SEC/DIV setting
Linearity over any 2 of
output
limit
center 8 divisions
(%)
(div)
A
B
.2 s
.1 s
≤ 0.1
10 nS/D
3
.5 s
.2 s
≤ 0.1
20 nS/D
3
1
s
.5 s
≤ 0.1
50 nS/D
3
.1 S/D
2
s
1
s
≤ 0.1
3
.2 S/D
5
s
2
s
≤ 0.1
3
.5 S/D
10
s
5
s
≤ 0.1
3
1 S/D
20
s
10
s
≤ 0.1
3
2 S/D
50
s
20
s
≤ 0.1
3
5 S/D
50
s
≤ 0.1
3
.1 ms
10 S/D
≤ 0.1
3
.2 ms
.1 ms
20 S/D
≤ 0.1
3
.5 ms
.2 ms
50 S/D
≤ 0.1
3
1 ms
.5 ms
≤ 0.1
.1 mS/D
3
2 ms
1 ms
≤ 0.1
.2 mS/D
3
5 ms
2 ms
≤ 0.1
.5 mS/D
3
10 ms
5 ms
≤ 0.1
1 mS/D
3
20 ms
10 ms
≤ 0.1
2 mS/D
3
50 ms
20 ms
≤ 0.1
5 mS/D
3
.1 sec
50 ms
A ONLY
(3) Adjust TI ⇐POSITION⇒ control to align 1st time marker with 1st (extreme left)
vertical graticule line.
(4) Adjust R740 (fig. 1) for 1 time marker per division over center eight divisions
(R).
(5) Set TI HORIZONTAL MODE switch to B and adjust B INTENSITY control
for suitable viewing. Adjust TI ⇐POSITION⇒ control to align 1st time marker with 1st
vertical graticule line.
(6) Adjust R730 (fig. 1) for 1 time marker per division over center eight divisions
(R).
(7) Set TI HORIZONTAL MODE switch to A and pull X10 CAL control to out
position.
(9) Adjust TI ⇐POSITION⇒ control to align nearest time marker to 1st vertical
graticule line.
(10) Adjust R754 (fig. 1) for 1 time marker per division (R).
(11) Set TI A AND B SEC/DIV switches to .2 ms.
(13) Adjust TI ⇐POSITION⇒ control to position middle time marker to center
vertical graticule line.
(14) Push TI X10 CAL control to in position.