TB 9-6625-2346-35
(40) Position TI controls as listed in (a) through (d) below:
(a) UPPER FUNCTIONS IN LOWER FUNCTIONS OUT pushbutton to the
OUT position.
(b) CH 1 V pushbutton pressed.
(c) CH 1 VOLTS/DIV switch to 50 m.
(d) CH 1 AC GND DC switch to GND.
(41) Adjust R1802 (fig. 3) for a TI indication of .0000 V.
(42) Set TI CH 1 AC GND DC switch to DC.
(43) While holding in TI P-P AUTO pushbutton, push TI SGL SWP pushbutton.
(44) Connect calibrator OUTPUT to TI CH 1 or X & DMM input.
(45) Set calibrator for an output of 400 mV.
(46) Adjust R1922 (fig. 4) for a TI indication of .4000 V (R).
35