(47) Repeat technique of (45) and (46) above for calibrator outputs and TI switch
settings listed in table 28. If TI readout indication is not within limits specified in table 28,
perform b (49) through (54) below.
Calibrator output
Test instrument
Voltage
Frequency
VOLTS/DIV
Min
Max
400 mV
20 kHz
50 m
.3954
.4046
4.0 V
20 kHz
0.5
3.954
4.046
40 V
20 kHz
5
39.54
40.46
(48) Set all outputs to minimum and disconnect equipment setup.
b. Adjustments
(1) Set all outputs to minimum and disconnect equipment setup.
(2) Press TI UPPER FUNCTIONS IN LOWER FUNCTIONS OUT pushbutton to
the IN position.
(3) Press TI PER pushbutton.
(6) Adjust C1311 (2236/2236A) (fig. 3) or TCXO CAL (Option 14) (fig. 3) for a TI
indication of 1.00000 s.
33