TB 9-6625-2346-35
(6) TI readout should indicate between 999.9900 kHz and 1.000010 MHz (999.9995
kHz and 1.000000 MHz for option 14). If not, perform b (1) through (9) below.
(7) Press TI PER pushbutton.
(8) TI readout will indicate between .99800 s and 1.00200 s.
(9) Disconnect equipment setup.
(10) Press TI DCV pushbutton.
(11) Connect a short between TI MULTIMETER INPUTS.
(12) TI readout will indicate between -.0001 and .0001 V, if not, perform b (10)
through (17) below.
(14) Connect calibrator OUTPUT terminals to TI MULTIMETER INPUTS.
(15) Set calibrator for a DC output of 400 mV. If TI readout indication is not within
limits specified in first row of table 24, perform b (18) through (23) below.
(16) Repeat technique of (15) above for calibrator outputs listed in table 24. If TI
readout indication is not within limits specified in table 24, perform b (18) through (23)
below.
Calibrator output
Test instrument indication limits
DC Volts
Min
Max
400 mV
.3995
.4005
-400 mV
-.3995
-.4005
4.0 V
3.995
4.005
-4.0 V
-3.995
-4.005
40 V
39.95
40.05
-40 V
-39.95
-40.05
400 V
399.5
400.5
-400 V
-399.5
-400.5
(17) Reduce calibrator output to minimum and disconnect equipment setup.
(18) Press TI UPPER FUNCTIONS IN LOWER FUNCTIONS OUT pushbutton to
the OUT position.
(19) Press TI AC RMSV pushbutton.
(20) Connect a short between TI MULTIMETER INPUTS.
(21) TI readout will indicate < .0006 V, if not, perform b (24) through (29) below.
(23) Connect calibrator OUTPUT terminals to TI MULTIMETER INPUTS.
(24) Set calibrator for an output of 400 mV at 20 Hz. If TI readout indication is not
within limits specified in first row of table 25, perform b (30) through (39) below.
30