TB 9-6625-2346-35
Calibrator
Nominal output
Error display limits
(Ω)
(%)
10
2.3
100
.4
1k
.4
10 k
.4
100 k 1
.4
1M
.4
10 M
.4
100 M
1.1
Set calibrator 2-WIRE COMP to OFF
1
(36) Disconnect equipment setup.
(37) Position TI controls as listed in (a) through (d) below:
(a) UPPER FUNCTIONS IN LOWER FUNCTIONS OUT pushbutton to the
OUT position.
(b) CH 1 V pushbutton pressed.
(c) CH 1 VOLTS/DIV switch to 50m.
(d) CH 1 AC GND DC switch to GND.
(38) TI readout will indicate between -.0012 and .0012, if not, perform b (40) and (41)
below.
(39) Set TI AC GND DC switch to DC.
(40) Connect calibrator OUTPUT to TI CH 1 or X & DMM input.
(41) Set calibrator for an output of 400 mV. If TI readout indication is not within
limits specified in first row of table 27, perform b (42) through (48) below.
(42) Repeat technique of (37) (c) and (41) above for calibrator outputs and TI switch
settings listed in table 27. If TI readout indication is not within limits specified in table 27,
perform b (42) through (48) below.
Calibrator
Test instrument
Output (DC Volts)
VOLTS/DIV
Indication limits
Min
Max
400 mV
50 m
.3982
.4018
4.0 V
0.5
3.982
4.018
40 V
5
39.82
40.18
(43) Set calibrator output to minimum.
(44) Set TI AC GND DC switch to AC.
(45) Set TI CH 1 VOLTS/DIV switch to 50m.
(46) Set calibrator for an output of 400 mV at a frequency of 20 kHz. If TI readout
indication is not within limits specified in first row of table 28, perform b (49) through (54)
below.