TB 9-6625-2365-35
(h) Enter ([Off] selected).
(i) Return to Meas.
(j) Resolution > (to 0.001Hz resolution).
amplitude controls for 25 mV rms output. TI will indicate within limits specified in first
(4) Repeat (3) above for remaining frequencies listed in table 3. TI will indicate
Table 3. Input 2
Sensitivity (10 Hz to 10 MHz)
Test instrument indication limits
Amplitude (rms)
Min
Max
10
Hz
25 mV
9.999
10.001
100
Hz
25 mV
99.999
100.001
1
kHz
25 mV
999.999
1000.001
10
kHz
25 mV
9999.999
10000.001
100
kHz
25 mV
99999.999
100000.001
1
MHz
25 mV
999999.999
1000000.001
10
MHz
25 mV
9999999.999
10000000.001
7