TB 9-6625-2365-35
Figure 3. Input 1
sensitivity (600 MHz to 18
GHz).
(4) Repeat (3) above for synthesized signal generator frequencies and power meter
no. 1 indications listed in table 5. TI will indicate within limits specified in table 5.
Table 5. Input 1
Sensitivity (600 MHz to 18
GHz)
Synthesized signal
Power meter
Test instrument indication limits
no. 1
(Hz)
Indication
(dBm)
Min
Max
600 MHz
-33.0
599999999.9
600000000.1
1 GHz
-33.0
999999999.9
1000000000.1
5 GHz
-33.0
4999999999.8
5000000000.2
10 GHz
-33.0
9999999999.8
10000000000.2
12.5 GHz
-28.0
12499999999.8
12500000000.2
15 GHz
-28.0
14999999999.7
15000000000.3
18 GHz
-28.0
17999999999.7
18000000000.3
(5) Reduce all outputs to minimum and connect equipment as shown in figure 4
connection A.
CHANGE 2