Figure 4. Stimulus Power Output Level Setup.

Calibration Procedure For Hand-Held Radio Test Set An/Prm-36 Page Navigation
  3    4    5    6    7  8
TB 9-6625-335-24
(2) Press the spectrum analyzer keys as listed in (a) through (g) below.
(a)
MODE, [Spectrum Analysis]
(b)
PRESET
(c)
FREQUENCY, [Center Freq], 100 [MHz]
(d)
SPAN, 1.1 [MHz]
(e)
AMPLITUDE, [Ref Level], -80 [dBm]
(f)
BW/AVG, [Average], [On] 100, [Enter]
(g)
SWEEP, [Sweep] (select single)
(3) Toggle SEL button ("STIM" will appear in the upper left corner of TI display).
(4) Press RESTART button on spectrum analyzer.
NOTE
100 measurements will accumulate within 5 seconds time. This
accumulation is visible on the left side of spectrum analyzer
display right below "LgAv".
(5) After average number reaches 100, press PEAK SEARCH on spectrum analyzer.
(6) Marker will indicate between -100.000 and -94.000 dBm.
(7) Disconnect equipment.
b. Adjustments. No adjustment can be made.
11. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/blank