TB 9-6625-2296-35
(9) Measuring receiver FM indication will be <140 Hz.
(10) Enter 9 9 9 M using TI DATA ENTRY keys.
(11) Measuring receiver FM indication will be <140 Hz.
(12) Set measuring receiver to measure AM.
(13) Enter 1 M using TI DATA ENTRY keys.
(14) Measuring receiver AM indication will be <0.10%.
(15) Step TI frequency in 100 MHz increments from 100 MHz to 999.9999 MHz using
technique of (13) above. Measuring receiver AM indications will be <0.10%.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a)
MODE MTRS ..........................
Auxiliary Functions Menu
(b)
AUX (F6) ..................................
Recall Parameters Menu
(c)
MEMORY RCL .......................
FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
(d)
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ........... Factory Defaults restored
(2) Connect frequency counter A input to TI T/R IN/OUT.
(3) Press key sequence listed in (a) through (h) below:
RF Gen Display
(a)
MODE RF GEN ......................
Gen Menu
(b)
SETUP .....................................
RF Gen Setup
(c)
DATA ENTRY 5 ......................
RF Gen Level
(d)
DATA ENTRY 2 ......................
0 dBm
(e)
DATA ENTRY 0 ENTER .......
RF Gen Freq
(f)
DATA ENTRY 1 ......................
500 kHz
(g)
DATA ENTRY 5 0 0 K ............
Return
(h)
Ret (F5) ....................................
(4) Frequency counter will indicate within limits listed in table 10.
(5) Press TI DATA SCROLL arrow keys to set MODE RF GEN frequency to settings
listed in table 10. Frequency counter indications will be within limits specified in table 10.