TB 9-6625-2296-35
b.
Adjustments
(1) Press key sequence listed in (a) through (l) below:
Meter Menu
(a) MODE MTRS ..........................
Auxiliary Functions Menu
(b) AUX (F6) ..................................
Recall
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER............
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER............
RF Gen Display
(g) MODE RF GEN ......................
Gen Menu
(h) SETUP .....................................
RF Gen Setup
(i) DATA ENTRY 5 ......................
RF Gen Level
(j)
DATA ENTRY 2 ......................
0 dBm
(k) DATA ENTRY 0 ENTER .......
Escape
(l) ESC (F6) ...................................
(2) Press key sequence listed in (a) through (d) below:
Auxiliary Function Menu
(a) AUX (F6) ..................................
(b) DATA ENTRY 1 ......................
(c) FIELD SELECT ⇑ ..................
Hightlight 14. Gen Output Level
Select Option
(d) DATA ENTRY ENTER twice
(3) Use (F1) key to toggle between "HIGH" and "LOW" ends of each range of
frequencies and record the two levels appearing on measuring receiver. Algebraically
subtract the two levels of each frequency range to verify levels are balanced. If not, adjust
DATA SCROLL knob and press arrow keys to adjust for a balance, then press DATA
ENTRY ENTER key to store new cal factors. Verify new cal factors by toggling between
"HIGH" and "LOW" for each frequency range adjusted.
(4) Press FIELD SELECT arrow key to select the next option (example: 2. 500 kHz) and
repeat (3) above.
(5) Repeat (4) above for each Gen Level Calibration option shown on TI.
(6) Press MODE RF GEN (RF GEN Display) key; when prompted for backup data,
press SHIFT and Y keys (Backup Cal Data ? y/n).
(7) Press ESC (F6) key.
24