TB 9-6625-2296-35
Low Pass Filter Freq 30.000 kHz
(d) DATA ENTRY 3 0 K ...............
Return to SINAD METER
(e) Ret (F5) ....................................
Use calibrator amplitude output
adjustment controls to set calibrator control display Error to +96.242%. TI SINAD meter
will indicate between 27.4 and 29.6 dB.
b. Adjustments
(1) Press key sequence listed in (a) through (d) below:
SINAD Meter Menu
(a) SETUP .....................................
Auxiliary Functions Menu
(b) AUX (F6) ..................................
(c) DATA ENTRY 1 ......................
Distortion Meter
(d) DATA ENTRY 6 ......................
(2) Press DATA ENTRY ENTER key to highlight Proper Reading field. Enter
10 using DATA ENTRY keys and press DATA ENTRY ENTER key. Press Ret (F5) key.
(3) Press MODE MTRS (Meter Menu) key; when prompted for backup data press
SHIFT and Y keys (Backup Cal Data ? y/n).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI DMM V/Ω and COM.
(2) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS ..........................
Auxiliary Functions Menu
(b) AUX (F6) ..................................
Recall Parameters Menu
(c) MEMORY RCL .......................
(d) FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ..........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ..........
(3) Press key sequence listed in (a) through (c) below:
Meters Menu
(a) MODE MTRS ..........................
(b) FIELD SELECT ⇑ ..................
(c) DATA ENTRY ENTER ..........