TB 9-6625-2308-40
(3) Press MENU key.
(4) Scroll to CALIBRATOR and press ENTER key.
(5) Select EEPROM and press ENTER key.
(6) Press ENTER key to get past SNUMB (unit serial number).
(7) Enter CALFAC change determined in (2) above.
(8) Press ENTER key.
(9) Correct DATE and press ENTER key.
(10) Correct TIME and press ENTER key.
(11) Select WRITE and then press ENTER key.
(12) Enter correct password if needed. (Password may be cleared or set.)
(13) Press ENTER key.
(14) If jumper A2W1 on analog PC board had to be moved to position A in step (1)
above, return jumper A2W1 to position B.
a. Deenergize and disconnect all equipment and reinstall protective covers on TI
if necessary.
b. Annotate and affix DA label 163 (Linearity not Calibrated) in accordance with
TB 750-25.
POWER SENSOR, GIGA-TRONICS MODEL 80301A
10. Preliminary Instructions
a. The instructions outlined in paragraphs 10 and 11 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.