TB 9-4931-501-35
a. Performance Check
(1) Press TI POWER switch to off.
(2) Install calibration fixture (A2) in TI left compartment.
(3) Remove vertical amplifier (A6) from TI right compartment and remove time
base (A4) from center compartment.
(4) Install time base in TI right compartment and pull TI POWER switch to on.
(5) Position calibration fixture controls as listed in (a) through (d) below:
(a)
POSITION control to midrange.
(b)
AMPLITUDE control to midrange.
(c)
TEST switch to COM MODE.
(d)
REP RATE switch to 100 kHz (kilohertz).
(6) If trace is not within 0.5 division of the center horizontal graticule line,
perform b below.
(7) Press TI POWER switch to off and move calibration fixture from left to center
vertical compartment.
(8) Pull TI POWER switch to on. If trace is not within 0.5 division of center
horizontal graticule line, repeat (b) below for a compromise between left and center
vertical compartments.
b. Adjustments. Adjust VERT CENTER
R135 (fig. 4) to
position trace to
horizontal
graticule line.
a. Performance Check
(1) Press TI POWER switch to off and install calibration fixture (A2) in TI left
compartment.
(2) Pull TI POWER switch to on.
(3) Set calibration fixture TEST switch to VERT OR HORIZ GAIN and REP
RATE switch to 1 MHz.
10 CHANGE 6