Input Reflection
Final Procedure

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
TB 9-4931-538-40
(4) Tighten all the cable connections.
(5) Set POWER STBY/ON switch to STBY.
(6) Press and hold down the top or bottom vertical FUNCTION key, and set
POWER STBY/ON switch to ON continuing to hold key until the graticule lines are displayed.
(7) Set the display menu and mainframe settings to meet the following conditions:
(a) Press Display key. Display Mode to Average is highlighted. NUMBER
OF AVERAGES to # Averages = 16. Screen to Single is highlighted.
(b) Press Channels key. Channel 1 display On is highlighted. Channel 2
through 4 display Off is highlighted.  VOLTS/DIV to Sensitivity = 5.000 mV/div.
OFFSET to Offset = 200.0 mV.
(c) Press Timebase key. TIME/DIV to Sweep Speed = 350 ps/div. DELAY
to Delay = 23.5000 ns.
(d) Press MORE and Network key.
Reflect/Trans/Cal to Reflect is
highlighted. Step & Chan 1 to On is highlighted.
(8) Press Timebase and DELAY keys. Adjust the delay so the second rising edge
of the signal (representing an open circuit) is 4 divisions left of center of display.
(9) Press Network key and press the Reflect/Trans/CAL key until CAL is highlighted.
(10) Press Reflect Cal key. Connect a short with an APC 3.5 (f) connector to the
semi-rigid cable, and press Reflect Cal key.
(11) Remove the short from the semi-rigid cable and connect a 50  termination to
the semi-rigid cable. Press Reflect Cal key.
(12) Remove the 50  termination from the semi-rigid cable and connect the semi-
rigid cable to Channel 2. Press SYSTEM CONTROL CLEAR DISPLAY pushbutton and
allow 16 averages to accumulate.
(13) Press More and Network keys.
Press Reflect/Trans/Cal key until
Reflect is highlighted.
(14) Set the NORMALIZED RISETIME to Filter Risetime = 30 ps. Press the
Normalize to Mem 1 key.
(15) Press SYSTEM CONTROL STOP and CLEAR DISPLAY pushbuttons.
(16) Press Cursor key until mem 1 is highlighted.
(17) Press Min & Max Reflect key. The absolute value of Rho minimum and Rho
maximum will be less than 5% ( <5%).
(18) Press SYSTEM CONTROL RUN pushbutton.
(19) Loosen the connection where the S and U cables connect together, and move the
semi-rigid cable from Channel 2 to Channel 3. Press SYSTEM CONTROL CLEAR
DISPLAY pushbutton and retighten the connection where the semi-rigid S and U cables
join together.