TDR System - Continued
Input Reflection

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  16    17    18    19    20  21  22    23    24    25    26  
img
TB 9-4931-538-40
(21) Press SYSTEM CONTROL CLEAR DISPLAY pushbutton and allow 16
averages to accumulate.
(22) Press MARKER2 POSITION key and adjust marker to the highest point of
signal on display. Record the V reading. V=_____________mV.
(23) Divide the V recorded in (22) above by the value of (20) above and multiply by
100%. V reading ________________mV step (20) _________mV) x 100% = ____________ %.
The absolute value of this number should be less than 1 % (+ 1 % at < 1 ns after edge).
(24) Adjust MARKER2 to the lowest point of signal on display.

(25) Press Timebase and TIME/DIV keys.  Press front panel ENTRY
increment pushbutton to select the next fastest sweep speed.
NOTE
If there is a point on the display lower than the previous
position of MARKER2 POSITION, readjust MARKER2
POSITION to that lowest point.
(26) Repeat (25) above until Sweep Speed = 1.00 ns/div.
(27) Divide the value of V obtained from display by the value obtained in (20) above.
V__________mV + step (20) _____________mV x 100% =_______________ %. The absolute
value will be <1 % (-l % at <1 ns after edge).
(28) Press Timebase key and set Sweep Speed = 200 ps/div, Delay Ref at
Center is highlighted and DELAY to Delay = the value of the START MARKER.
(29) Press Channels key and set VOLTS/DIV to Sensitivity = 80.00 mV/div.
(30) Press Delta V key. Press Preset Level key until 50-50% is highlighted. Press
Auto Level Set key.
(31) Press Delta t key and press START ON EDGE key until POS 1 is highlighted.
Press STOP ON EDGE key until POS 1 is highlighted. Press Precise Edge Find key.
(32) Press the START and STOP MARKER keys and record start and stop
positions. START MARKER _________ns, STOP MARKER ___________ns.
(33) Press Timebase key and set Delay Ref at Left is highlighted. Set DELAY to
Delay = the value of START MARKER.
(34) Press Delta t key and adjust STOP MARKER to the value of the START
MARKER plus 500 ps.
(35) Press Timebase key and set Sweep Speed = 100 ps/div. Press Channels key
and set Sensitivity = 2.000 mV/div.
(36) Press Delta V key and adjust MARKER2 to the lowest point of signal
displayed. DO NOT INCLUDE the rising edge of the signal. If the Delta V point is to the
left of the STOP MARKER, subtract the value obtained in (20) above from V(2) on display.
Record this value as in (37) below. V(2)=_____________mV.