• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Figure 7. Trigger sensitivity low frequency, 100 MHz and 2.5 GHz - equipment setup.
Figure 8. Trigger sensitivity jitter - equipment setup.

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. .
TB 9-4931-538-40
(12) Set signal generator for a frequency of 100 MHz and amplitude of V p-p 150 mV
as displayed on screen.
(13) Slowly adjust the RPG control cw until trigger level just stops triggering.
Record the trigger level setting (_____________). This value is Vpos.
(14) Reset the trigger level to 0.000 V and slowly adjust the RPG control ccw until
trigger level just stops triggering. Record the trigger level setting (_____________). This
value is Vneg.
(15) Calculate the sensitivity. Vsense = 150 mV - (Vpos - Vneg). The value will be
40 mV.
(16) Reset the trigger level to 0.000 V.
(17) Set HF Reject to On is highlighted and repeat (13) through (15) above. The
value will be 57 mV (+ 3 dB).
(18) Reset the trigger level to 0.000 V.
(19) Press Timebase key and set Sweep Speed = 100 ps/div.
(20) Press Channels and VOLTS/DIV keys and set Sensitivity = 50.00 mV/div.
(21) Press Trigger key and set TRIGGER LEVEL to Trigger Level = 0.000 V; HF
Sens and HF Reject Off is highlighted.
(22) Connect equipment as shown in figure 7, CONNECTION C.
(23) Set synthesized signal generator for a frequency of 2.5 GHz and amplitude of
V p-p 200 mV as shown on display.
NOTE
It will be necessary to go to the Measure Mode to verify the
measurement of the V p-p = 200 mV.
(24) Slowly adjust the RPG control cw until the trigger level just stops triggering.
Record the trigger level setting (_____________). This value is Vpos.
(25) Reset the trigger level to 0.000 v and slowly adjust the RPG control ccw until
trigger level just stops triggering. Record the trigger level setting (_____________). This
value is Vneg.
(26) Calculate the sensitivity. Vsense = 200 mV - (Vpos - Vneg). The value will be
<200 mV.
(27) Set Trigger Level = 0.000V. Set delay to 35 ns.
(28) Connect equipment as shown in figure 8.
17