Table 5. Short Delta t
Table 6. Long Delta t

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . .
TB 9-4931-538-40
Figure 5. Time interval short delta t - equipment setup.
(11) Press SYSTEM CONTROL/ RECALL SETUP and ENTRY 3 pushbuttons.
Press Channel key and set Channel 1 to Display On is highlighted.
(12) Connect equipment as shown in figure 6.
Figure 6. Time interval long delta t - equipment setup.
(13) Set function generator for a frequency of 10 MHz sine wave and amplitude of 1.2 V p-p.
(14) Press More, Timebase, and TIME/ DIV keys and set Sweep Speed = 20.0 ns/div.
(15) Press Trigger key and set TRIGGER LEVEL to Trigger Level = 500 mV.