• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Figure 11. Horizontal assembly A1 adjustments.
Frequency Adjustment

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
img
. . . .
TB 9-4931-538-40
Figure 12. Vertical assembly A2 adjustments.
e. Readjust the channel offset to re-center the signal on display.
f. Press Delta V key. Set both V MARKERS ON and overlay them on Channel 1.
g. Press Preset Levels key until Preset Levels 0-100% is highlighted.
h. Press Channel and OFFSET keys and adjust RPG control to center signal on display.
i. Press Delta V and Auto Level Set keys.
j.  Press Display key. Press Display Mode key until Persist is highlighted and set
Persistence = 300 ms.
k. Preset SRD BIAS on horizontal assembly A1 (fig. 11) to mechanical center.
l.  Adjust SRD DRIVE on horizontal assembly A1 (fig. 11) for a signal amplitude of 0%
and + 100% on display.
NOTE
An offset may exist in the signal during this portion of the
adjustments. If the signal peaks do not overlay the 0% and
100% markers, adjust SRD DRIVE (fig. 11) for equal amounts
of signal above or below the markers in the following steps.
The OFFSET level may be adjusted anytime during this
procedure.